beautypg.com

Campbell Scientific CS135 Lidar Ceilometer User Manual

Page 33

background image

User Guide

25

command)

Line

Example line output

8

Heaters A B C

Description of the line sections
Section Description
A

Hood blower mode

B Internal

heater

mode

C

Laser heater mode

Line

Example line output

9

MCFG X A B C D E

Description of the line sections
Section Description
X Message

interval

(Note: Refer to the MCFG command)

A

Message ID A (Note: Refer to the MCFG command)

B

Message ID B (Note: Refer to the MCFG command)

C

Message ID C (Note: Refer to the MCFG command)

D

Message ID D (Note: Refer to the MCFG command)

E

Message ID E (Note: Refer to the MCFG command)

Line

Example line output

10

Angle A B C

Description of the line sections
Section Description
A X

axis

tilt

B Y

axis

tilt

C

Beam angle from vertical

Line

Example line output

11 Units

A

Description of the line sections
Section Description
A

Measurement units and tilt correction. (Note: Refer to the UNITS command)

Line

Example line output

12

TRH A B C

Description of the line sections
Section Description
A

Sensor internal temperature reading in degrees Celsius.

B

Sensor internal humidity reading as a percentage

C

Sensors internal dew point value in degrees Celsius

Line

Example line output

13

T A B C D E F

Description of the line sections
Section Description
A

External fan blower assembly temperature in degrees Celsius

B

PSU internal temperature in degrees Celsius

C

TOP board laser monitor temperature in degrees Celsius

D

TOP board calibration LED temperature in degrees Celsius

E

Laser module temperature in degrees Celsius

F

Photo diode module temperature in degrees Celsius

Line

Example line output

14

SupplyVoltage A B

Description of the line sections
Section Description
A

DSP board supply voltage

B