CS135 Ceilometer

22

Message_ID_D
Message_ID_E

Note that if low baud rates and message intervals or short polling are
required it may not be possible to output multiple longer messages.

Message_ID is the message type to output between 0 and 999 (default
001). If Message_ID = 0 no message type will be output. Up to 5
messages can be set. Please refer to Section 5.1.4.

NEWLASER

No parameters

Resets the laser run time counter.

OPEN

ID
Password

Open the CS135 terminal mode.

ID = Sensor ID as per the terminal command “ID”

Password = The sensors user password as per the terminal command
“PASSWORD”. The default is no password.

PASSWORD

Password

Sets or clears a password from 1 to 10 characters in length. Valid ASCII
characters 0-9, a - z or A – Z and letters are case sensitive. Typing the
command PASSWORD without any parameters clears the password.
The default is no password.

POLL

Sensor_ID
Message_ID

Requests the message Message_ID from the sensor Sensor_ID.
Refer to the Section 5.1.6 for more information on this command
Note: Message_ID outputs the message configured by MCFG.

PSU

Blower_Voltage
Boost_Volts
Battery_Current
PSU_Present

Sets PSU battery charge parameters and external blower speed.

Blower_Voltage sets the voltage applied to the external blower in milli-
volts (mV). From 8000 to 14000. The default is 12000.

Boost_Volts sets the boost voltage to be applied to the PSUs battery in
mV. From 11000 to 16000. (Note: It is not recommended that the
customer alters this parameter without first consulting Campbell
Scientific.). The default is 14520.

Battery_Current sets the charge current in milli-amps (mA). From 100
to 1500. The default is 400.

PSU_Present sets a flag indicating if a PS135E PSU is connected to the
sensor.
PSU_Present = 0, Power supply is not fitted.
PSU_Present = 1, Power supply is fitted. (default)

REBOOT

No parameters

Forces a system reboot. This will restore previously saved user settings.
Any unsaved changes will be lost. (Settings are saved in terminal mode
when the command CLOSE is typed, which exits the terminal mode).

SENSITIVITY

Boundary
Detector
Alpha

Boundary From 1 to 20, default 4
Detector From 1 to 20, default 6
Alpha From 0.0001 to 0.1, default 0.01
(Do not change without consulting Campbell Scientific Ltd)

SERIAL

Mode
Baud
Bits_Parity
Delay

Set or read the serial port

Mode = 0, RS232, full duplex (default).
Mode = 1, RS232, half duplex.
Mode = 2, RS485, full duplex.
Mode = 3, RS485, half duplex.
Mode = 4, reserved.
Mode = 5, RS422.

Baud = 0, 300 baud.
Baud = 1, 600 baud.
Baud = 2, 1200 baud.
Baud =3, 2400 baud.
Baud = 4, 4800 baud.
Baud = 5, 9600 baud.