Unit, Table 24, Form command modifiers – Vaisala MMT330 User Manual
Page 99
Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 99
- Water activity (aw)
- Temperature (T) (metric unit: ºC, non-metric unit:ºF)
When selecting the quantity, use the abbreviations of the quantities.
The modifiers are presented in Table 24 on page 99 below.
Table 24
FORM Command Modifiers
Modifier
Description
x.y
Length modifier (number of digits and decimal places)
#t
Tabulator
#r
Carriage-return
#n
Line feed
“”
String constant
#xxx
Special character, code "xxx" (decimal), for example #027
for ESC
U5
Unit field and length (length optional)
ADDR
Transmitter address [00...255]
SN
Transmitter serial number
TIME
Time [hh:mm:ss]
DATE
Date [yyyy-mm-dd]
Example:
>form "aw=" 6.4 aw #t "t=" 6.2 t #r#n
OK
>send
aw= 0.2644
t= 25.50
>
Command 'FORM /' will return the default output format. The default
output format depends on the device configuration.
>form /
OK
>send
aw= 0.087 T= 24.0 'C
>
UNIT
Use the UNIT command to select metric or non-metric output units:
UNIT [x]
where
x
=
M or N
where
M
=
Metric units
N
=
Non-metric units