Feeding Adjustment Information
This information is shown on the device information fields (see section
Device Information on page 102.)
Using Display/Keypad
1.
If you are not in the adjustment menu, press the ADJ button on the
motherboard (opens the ADJUSTMENT MENU).
2.
Select ►Adjustment info, press the arrow ► button.
3.
Select Date, press SET. Give the date by using the arrow buttons.
Press OK.
4.
Select i, press SET. Enter information text including 17 characters
at maximum by using the arrow buttons. Press OK.
5.
Press EXIT to return to the basic display.
Using Serial Line
CTEXT
Use the CTEXT command to enter text to the adjustment information
field.
Example:
>ctext
Adjust. info : (not set) ? HMK15
>
CDATE
Use the CDATE command to enter date to adjustment information field.
Set the adjustment date in format YYYY-MM-DD.
Example:
>cdate
Adjust. date : (not set) ? 2004-05-21
>