beautypg.com

Stopping continuous outputting, Outputting reading once, Assign an alias for the send command – Vaisala MMT330 User Manual

Page 95

background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 95

- Outputting interval can be changed with the INTV command.

- Output message format can be changed with the FORM command.

Stopping Continuous Outputting

Use the S command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.

S

See command SMODE to change the default (power-up) operation
mode.

Outputting Reading Once

Use the SEND command to output the reading once in STOP mode.

SEND

The output format depends on which parameters the transmitter can
output.

Examples:

>send
aw= 0.261 T= 23.8 'C
>

Assign an Alias for the SEND Command

Use the SCOM command to assign a new command that works like the
SEND command. The standard SEND command of the transmitter will
always function normally whatever the SCOM definition may be.
Command names are case-insensitive.

SCOM

Example (assigning the command MEASURE as an alias for the SEND
command):

>scom
Send command : ? measure
>measure
aw= 0.261 T= 23.8 'C
>

You can remove the SCOM definition by pressing Esc when entering the
command name:

>scom
Send command : measure ?