Table 29

Error Messages

Error Code Error Message

Action

E0

Humidity sensor measurement
malfunction.

Check the integrity of the humidity probe
and the probe cable. Clean the probe from
dirt, water, ice or other contaminants.

E1

Humidity sensor short circuit

E2

Humidity sensor open circuit

Check the integrity of the humidity probe
and the probe cable.

E3

Temperature sensor open
circuit.

E4

Temperature sensor short
circuit.

Check the integrity of the humidity probe
and the probe cable. Clean the probe from
dirt water, ice or other contaminants.

E5

Temperature measurement
malfunction

E6

Temperature sensor current
leak.

Check the integrity of the humidity probe
and the probe cables. Clean the probes
from dirt, water, ice or other contaminants.

E7

Internal ADC read error

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E9

Checksum error in the internal
configuration memory

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E10

Internal EEPROM read error

E11

Internal EEPROM write error

E12 … E13

Add-on module 1 (or 2)
connection failure

Turn off the power and check the module
connection. Turn on the power.

E14

Device internal temperature
out of range

Ensure that the operating temperature is
within the valid range.

E15

Internal RAM memory or
software checksum error

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E18

Internal ADC reference voltage
out of range

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E19

Internal analog output
reference voltage out of range

E20 … E22

Configuration switches for
analog output 1/2/3 set
incorrectly

Check and re-set the switches, see Figure
58
on page 114 and Figure 30 on page 53.

E24 … E25

Internal error in add-on module
1 (or 2)

Disconnect the power and check the
module connection.

E26

Communication module
installed in incorrect add-on
module slot

Disconnect the power and change the
communication module to another module
slot.

E28 … E29

Unknown/incompatible module
installed in add-on module slot
1(or 2)

Ensure that the module is compatible with
the MMT330.

E30

Internal analog voltage out of
range

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E31

Internal system voltage out of
range

Check that the supply voltage is correct
and the power supply can provide enough
power for the instrument.