beautypg.com

Analog output adjustment, Using display/keypad, Using serial line – Vaisala MMT330 User Manual

Page 149: Acal

background image

Chapter 7 ___________________________________________________ Calibration and Adjustment

VAISALA _______________________________________________________________________ 149

Analog Output Adjustment

In the analog output calibration the analog output is forced to the
following values:

- Current output: 2 mA and 18 mA

- Voltage output: 10 % and 90 % of the range

Connect MMT330 to a calibrated current/voltage meter in order to
measure either current or voltage depending on the selected output type.

Using Display/Keypad

1.

Press the ADJ button to open the ADJUSTMENT MENU.

2.

Select ►Adjust analog outputs, press ►key.

3.

Select the output to be adjusted Adjust analog output 1/2, press
START.

4.

Measure the first analog output value with a multimeter. Give the
measured value by using the arrow buttons. Press OK.

5.

Measure the second analog output value with a multimeter. Give
the measured value by using the arrow buttons. Press OK.

6.

Press OK to return to the adjustment menu.

7.

Press EXIT to close the adjustment mode and to return to the basic
display.

Using Serial Line

Enter the ACAL command and type the multimeter reading for each
case. Continue by pressing ENTER.

ACAL

Example (current outputs):


>acal
Ch1

I1

(mA) ?

2.046

Ch1

I2

(mA) ?

18.087

Ch2

I1

(mA) ?

2.036

Ch2

I2

(mA) ?

18.071

>