Example (set aw and T as display alarm quantities and configure the
alarms as prompted):

>alsel aw t
Al1 aw above: 0.50 ? 0.6
Al1 aw below: - ?
Al1 aw hyst : 0.10 ?
Al1 aw enabl: ON ?
Al2 T above: - 'C ?
Al2 T below: - 'C ? 0
Al2 T hyst : 0.00 'C ? 1
Al2 T enabl: OFF ? ON
>

MI70 Link Program for Data Handling

MI70 Link is a support program that enables you to handle the
measurement data of your Vaisala transmitter using a Windows PC.
Using the MI70 Link you can, for example:

- Monitor the transmitter’s measurement readings directly using the

real-time window function.

- Transfer recorded data from the main transmitter memory in numeric

or graphical format, for further use in a spreadsheet program (such as
Microsoft

Excel) or virtually any other application. Note that the

maximum amount of data points that can be downloaded in one
transfer is 65000.

To use the MI70 Link program, you must connect your Windows PC to
the service port of the transmitter with a serial or USB cable. The MI70
Link program, and the optional connection cables, are available from
Vaisala. See section Spare Parts and Accessories on page 156.

Follow the steps below to connect your transmitter to the MI70 Link
program using the serial interface:

1.

Connect your PC to the service port of the transmitter. Refer to
section Service Port Connection on page 80.

2.

Check that the MMT330 is powered.

3.

Start the MI70 Link program.

4.

Start using the program. There is usually no need to select a COM
port manually, the MI70 Link software can detect it automatically.

NOTE