Figure 26, Wiring of the m12 (8-pin) connector, Table 11 – Vaisala PTB330TS User Manual
Page 67: M12 connector pin assignments
Chapter 6 ______________________________ Serial Line Communication With PTB330 Barometer
VAISALA_______________________________________________________________________ 65
To take the user port into use for a serial connection:
-
Connect to the service port first, and configure the user port
communication settings using the SMODE and SERI commands.
The new user port settings will be taken into use when the
transmitter is reset or powered up. For example:
SMODE STOP
SERI 9600 E 7 1
-
Connect to the M12 connector using a serial cable with the proper
wiring. The wiring of the M12 connector is described in
. Note that you also have to provide the operation power
for the barometer (10 ... 35 VDC) via the cable.
0503-026
Figure 26
Wiring of the M12 (8-Pin) Connector
Table 11
M12 Connector Pin Assignments
Pin/Terminal
Wire
RS-485 (EIA-485)
1
White
A -
2
Brown
Serial GND
3 (optional)
Green
Serial GND
4
Yellow
-
5
Grey
Supply -
6
Pink
Supply +
7
Blue
B -
8
Shield/Red
Cable shield