Chapter 4 ______________________________________________________________ Maintenance
VAISALA_______________________________________________________________________ 47
0705-091
Figure 21
Inside of Open Barometer
1-Point Adjustment (LCP1)
The LCP1 command performs linear correction for barometer
module/module P1. Use the command LCP1
-
to activate or deactivate the linear adjustment function
-
to enter new linear offset and offset/gain pressure corrections to the
transmitter
-
to edit existing linear offset and offset/gain pressure corrections.
Numbers refer to
:
1
=
Adjustment button with indicator LED
2
=
Power supply module (not included in PTB330TS)
3
=
Power supply mode selections (do not change)
4
=
Service port (RS-232)
5
=
Module 1/Module 3 connectors
6
=
User port (wired to the M12 connector)
7
=
Module 2/Module 4 connectors