Multipoint adjustment (mpcp1) – Vaisala PTB330TS User Manual
Page 50
User's Guide ______________________________________________________________________
48 ___________________________________________________________________M210983EN-A
Key in the linear corrections for each barometer module separately.
Feeding ON or OFF as a parameter enables or disables the linear
correction. A single question mark shows the reading and reference
information stored and the state of the linear correction. Each barometer
module has its own adjustment commands. If there are three barometer
modules installed, the commands LCP1, LCP2, and LCP3 will be
available. First deactivate the previous corrections by using the LCP1
OFF command.
LCP1 [x/y] [z]
Example
>lcp1
1. Reading ? 980.2
1. Reference ? 980.0
2. Reading ? 1100.1
2. Reference ? 1100.0
>
Multipoint Adjustment (MPCP1)
Use the MPCP1 command
-
to perform multipoint correction for barometer module/module P1
-
to activate or deactivate the multipoint adjustment function
-
to enter new linear offset and offset/gain pressure corrections to the
transmitter
-
to edit existing linear offset and offset/gain pressure corrections.
where
x
=
ON
y
=
OFF
z
=
?
NOTE
The new linear corrections will always cancel the previous corrections
as well as the valid date of calibration of the transmitter.