beautypg.com

Calibration and adjustment of ptb330, Figure 20, Ptb330 adjustments – Vaisala PTB330TS User Manual

Page 47: Calibration

background image

Chapter 4 ______________________________________________________________ Maintenance

VAISALA_______________________________________________________________________ 45

Calibration and Adjustment of PTB330

In PTB330, the raw measured pressure value of each barometer module
is first corrected using the corresponding multipoint adjustment values.
The additional linear adjustment (if enabled) is applied to the corrected
values. This allows the user to fine-tune the pressure reading using the
linear adjustment without performing a complete readjustment for the
barometer module, see

Figure 20 on page 45

.

0708-053

Figure 20

PTB330 Adjustments

To measure new adjustment values and perform the actual adjustment,
first disable the corresponding adjustment on each barometer module
you are going to adjust (see the procedure above). If you are going to
perform a multipoint adjustment, disable also the linear adjustments. If
you are going to do only a linear adjustment, leave the multipoint
adjustments untouched.

After disabling the previous adjustments, calibrate the PTB330 against
your pressure reference. Make a note of the pressure reading of each
PTB330 barometer module separately (quantities P

1

, P

2

, and P

3

). Then

use the display/keypad to enter and activate the new multipoint/linear
adjustment factors.

The user can select a simple offset or a two-point offset and gain
adjustment and use the LCP command for adjustment of the barometer
module. The MPCP command is used for the more sophisticated
multipoint correction capability at up to eight pressure levels. Check
first what linear corrections the transmitter is currently using before

NOTE

Making adjustments is possible only after adjustments have been
unlocked. To unlock the adjustment menu, press the ADJ button on the
motherboard of the transmitter.