User's Guide ______________________________________________________________________

42 ___________________________________________________________________M210983EN-A

Do not push the connector with conducting material, for example
with a metallic screwdriver.

0811-094

Figure 18

Changing the MI70 Battery Pack

5.

Position the battery pack in inside the case, replace the back plate,
and tighten the screw.

6.

Charge the battery pack before use. Refer to section

Charging on

page 29

.

Changing the HMP155 Probe Filter

New filters for the HMP155 probe can be ordered from Vaisala; see

Options and Accessories on page 85

. Change the filter as follows:

1.

Remove the filter from the probe by turning it counter-clockwise.
When pulling out the filter, be sure not to damage the sensor. Refer
to

Figure 19 on page 43

.

2.

After removing the filter, check the O-ring and replace it if
necessary.

3.

Install a new filter on the probe.