beautypg.com

Entering adjustment information, Calibration and adjustment of hmp155 – Vaisala PTB330TS User Manual

Page 52

background image

User's Guide ______________________________________________________________________

50 ___________________________________________________________________M210983EN-A

Entering Adjustment Information

Adjustment information and date are a part of the device information
that is shown using the ? command. Use the CTEXT and CDATE
commands to enter updated adjustment information.

CTEXT

Use the CTEXT command to enter text to the adjustment information
field.

Example:

>ctext Vaisala/MSL

Calibration text : Vaisala/MSL

>

CDATE

Use the CDATE command to enter date to adjustment information field.
Set the adjustment date in format YYYY-MM-DD.

>cdate 2006-06-12

Calibration date : 2006-06-12

>

Calibration and Adjustment of HMP155

The procedures below provide a simple way to check and adjust the
HMP155 using the MI70 indicator. You can perform a 1-point or
2-point adjustment on relative humidity and temperature.

The humidity quantity to be adjusted is always RH. The temperature
quantity that you can adjust using the MI70 indicator depends on your
hardware options, since the HMP155 can have two temperature sensors:

-

If you do not have the additional temperature probe, the quantity
you can calibrate and adjust is T. It is measured on the HMP155
probe, next to the HUMICAP® sensor.

-

If you have the additional temperature probe, the quantity is T

a

.

It is measured by the separate temperature probe that is attached to
the HMP155.