beautypg.com

6 technical data, Specifications, Technical data – Vaisala HMDW110 User Manual

Page 58

background image

6

TECHNICAL DATA

Specifications

Property

Specification

Relative humidity

Measurement range

0 ... 100 %RH

Accuracy

Temperature range +10 ... +30 °C (+50 ... +86 °F)

0 ... 90 %RH

±2 %RH

90 ... 100 %RH

±3 %RH

Temperature range -20 ... +10 °C, +30 ... +60 °C
(-4 ... +50 °F, +86 ... +140 °F)

0 ... 90 %RH

±3 %RH

90 ... 100 %RH

±4 %RH

Temperature range -40 ... -20 °C (-40 ... -4 °F)

0 ... 100%RH

±4 %RH

Stability in typical HVAC applications

±0.5 %RH/year

Humidity sensor

HUMICAP

®

180R

Temperature

Measurement range

-40 ... +60 °C (-40 ... +140 °F)

Accuracy at +20 °C (+68 °F)

±0.3 °C

Temperature dependence

±0.01 °C/°C

Temperature sensor

Pt1000 RTD Class F0.1 IEC 60751

Calculated parameters

Measurement range for dewpoint temperature and
wet bulb temperature

-40 ... +60 °C (-40 ... +140 °F)

Measurement range for enthalpy

-40 ... 460 kJ/kg
(-10 ... +190 BTU/lb)

Accuracy of the calculated parameters should be calculated at the actual condition based on
the RH and temperature specification.

Accuracy at 20°C (68°F) and 80 %RH

Dew point

±0.7 °C (1.2 °F)

Wet bulb temperature

±0.5 °C (0.9 °F)

Enthalpy

±1.6kJ/kg (0.7 BTU/lb)

Table 40 Performance

56

6 Technical Data