Measurement output format – Vaisala HMDW110 User Manual
Page 31

Syntax
Description
s
Stop the continuous outputting of
measurement values.
Example:
...
T= 22.8 'C RH= 39.5 %RH Td=
8.3 'C Tw= 14.5 'C h=
40.4 kJ/kg
T= 22.8 'C RH= 39.5 %RH Td=
8.3 'C Tw= 14.5 'C h=
40.4 kJ/kg
s
Table 12 S Command
Since the interface is half-duplex, you must enter the s command
when the device is not outputting.
Syntax
Description
intv
Show the output interval of the automatically
repeating measurement messages (r
command and run mode).
intv [iii u]
Set the output interval.
iii
= interval, range 0 ... 255.
u
= unit for interval setting:
n
s = seconds
n
m = minutes
n
h = hours
If you set the interval to 0, the output
messages are output as quickly as they are
generated, without additional delay.
Example:
intv 5 s
Output interval:
5 S
Table 13 Intv Command
Measurement Output Format
Syntax
Description
unit
Show current setting of the unit command.
unit [m|n]
Set types of units used on the serial line.
m
= metric units, for example, Celsius
n
= non-metric units, for example, Fahrenheit
Example (set units to non-metric):
unit n
Units
: Non metric
Table 14 Unit Command
29
4 Service Port