Vaisala HMDW110 User Manual
Page 37

Syntax
Description
asel ?
Show currently set analog output
parameters and scaling.
asel
Show currently set analog output
parameters and scaling, prompt to enter new
scaling values.
asel [ch1 ch2]
Set new output parameters for both
channels, prompt to enter new scaling
values. Selectable parameters for ch1 and
ch2
are:
t
= temperature
rh
= relative humidity
td
= dewpoint
h
= enthalpy
tw
= wet bulb temperature
asel [ch1 ch2
[ch1lo ch1hi ch2lo ch2hi]]
Set analog output parameters and scaling for
both channels without prompting.
ch1
= Output parameter for channel 1.
Parameters as above.
ch2
= Output parameter for channel 2.
Parameters as above.
ch1lo
= Low limit for channel 1 scaling.
ch1hi
= High limit for channel 1 scaling.
ch2lo
= Low limit for channel 2 scaling.
ch2hi
=High limit for channel 2 scaling.
Example (show current analog output parameters and scaling):
asel ?
Ch1 RH
lo
:
0.00 %RH
Ch1 RH
hi
:
100.00 %RH
Ch2 T
lo
:
-40.00 'C
Ch2 T
hi
:
60.00 'C
Table 27 Asel Command
35
4 Service Port