beautypg.com

Calibration and adjustment commands – Vaisala HMDW110 User Manual

Page 38

background image

Syntax

Description

atest [ch1 ch2]

Set analog channels to defined output value
(in mA).

ch1

= Output value for channel 1.

ch2

= Output value for channel 2.

atest

End analog output testing mode, return
outputs to normal.

Example (set both channels to 20 mA, then return them to normal measurement):

atest 20 20
20.000

26393

20.000

26393

atest
6.694

26393

12.297

26393

Table 28 Atest Command

Calibration and Adjustment Commands

Before using the calibration and adjustment commands, read
through section Calibration and Adjustment on page 41.

Syntax

Description

crh

Start the two-point humidity calibration and
adjustment sequence. For a full adjustment
procedure, see section Two-Point Humidity
Calibration and Adjustment using a computer
and a HMK15 Humidity Calibrator on
page 46.

crh [ref]

Perform a one-point adjustment at the
current relative humidity. The device
calculates the required correction based on
the reference humidity that you enter.

ref

= the correct relative humidity at the

environment where the sensor is now.

Example :

crh

RH

:

11.5379 1. ref ?

11.3

Press any key when ready ...
RH

:

74.9684 2. ref ?

75.4

OK

Table 29 Crh Command

36

4 Service Port