Vaisala HMDW110 User Manual
Page 39

Syntax
Description
crhclr
Clear the current user adjustment for
humidity. Factory calibration remains.
Example:
crhclr
OK
Table 30 Crhclr Command
Syntax
Description
ct
Start the two-point temperature calibration
and adjustment sequence. For a full
adjustment procedure, see section Two-
Point Temperature Calibration and
Adjustment using a Computer on page 49.
ct [ref]
Perform a one-point adjustment at the
current temperature. The device calculates
the required correction based on the
reference temperature that you enter.
ref
= the correct temperature (in degrees
Celcius) at the environment where the
sensor is now.
Example :
ct
T
:
22.9424 1. ref ?
23.0
Press any key when ready ...
T
:
54.9873 2. ref ?
55
OK
Table 31 Ct Command
Syntax
Description
ctclr
Clear the current user adjustment for
temperature. Factory calibration remains.
Example:
ctclr
OK
Table 32 Ctclr Command
37
4 Service Port