Syntax

Description

crhclr

Clear the current user adjustment for
humidity. Factory calibration remains.

Example:

crhclr
OK

Table 30 Crhclr Command

Syntax

Description

ct

Start the two-point temperature calibration
and adjustment sequence. For a full
adjustment procedure, see section Two-
Point Temperature Calibration and
Adjustment using a Computer on page 49
.

ct [ref]

Perform a one-point adjustment at the
current temperature. The device calculates
the required correction based on the
reference temperature that you enter.

ref

= the correct temperature (in degrees

Celcius) at the environment where the
sensor is now.

Example :

ct

T

:

22.9424 1. ref ?

23.0

Press any key when ready ...
T

:

54.9873 2. ref ?

55

OK

Table 31 Ct Command

Syntax

Description

ctclr

Clear the current user adjustment for
temperature. Factory calibration remains.

Example:

ctclr
OK

Table 32 Ctclr Command

37

4 Service Port