Replacing the humicap – Vaisala HMDW110 User Manual

Page 53

background image

13. Give the l command to verify that new values for user adjustment

parameters T offset and T gain have been stored.

Example: user adjustment parameters after two-point temperature
adjustment. Your values will be different.

Cp offset :

 0.00000000E+00

Cp gain

:

1.00000000E+00

T

offset :

7.64255524E-02

T

gain

:

1.00767076E+00

14. Remove the probe from the temperature reference.

15. Disconnect the connection cable from the service port.

16. Reconnect the terminal blocks and close the cover.

Replacing the HUMICAP

®

 Sensor on HMD110/112

and HMW110/112

n

New HUMICAP

®

humidity sensor (HUMICAP180R or

HUMICAP180VHP).

n

New filter (always recommended when replacing the sensor).

n

Duct models only: medium size crosshead screwdriver
(Pozidriv).

1. Power down the transmitter.

2. Duct models only: loosen the tightening screw on the fastening flange, and

pull the probe out of the duct.

51

5 Maintenance