Device information and status – Vaisala HMDW110 User Manual

Page 27

background image

Command

Description

Page

crhclr

Clear user calibration for humidity measurement.
Factory calibration remains.

37

ct

User calibration for T measurement.

37

ctclr

Clear user calibration for temperature measurement. Factory
calibration remains.

37

fcrh

Two-point calibration after humidity sensor change.

38

l

Show adjustment offset and gain.

38

li

Set adjustment offset and gain.

39

Other commands

filt

Show or set measurement filtering.

39

frestore

Restore factory settings. Clears all user settings, factory
calibration remains.

40

reset

Reset the device.

40

Device Information and Status

Syntax

Description

?

Show listing of device information.

??

Show listing of device information even if
device is in poll mode and connection has not
been opened using the open command.

Example:

?
HMD112 / 1.0.6
Serial number

: H0134007

Batch number

: H0130002

Sensor number

: H0090003

Sensor model

: Humicap 180R

Order code

: HMD1102A1VA1

Cal. date

: 20140619

Cal. info

: VAISALA/HEL

Time

: 01:43:43

Serial mode

:

STOP

Baud P D S

:

19200 N 8 1

Output interval:

5 S

Serial delay

:

25

Address

:

0

Filter

:

1.000

Ch1 output

:

4 ... 20 mA

Ch2 output

:

4 ... 20 mA

Ch1 RH

lo

:

0.00 %RH

Ch1 RH

hi

:

100.00 %RH

Ch2 T

lo

:

-40.00 'C

Ch2 T

hi

:

60.00 'C

Table 5 ? Command

25

4 Service Port