Serial line output and communication, Measurement output – Vaisala HMDW110 User Manual
Page 30

Syntax
Description
time
Show transmitter uptime (time since last
reset) in hh:mm:ss.
Example:
time
Time
: 00:54:51
Table 9 Time Command
Serial Line Output and Communication
Measurement Output
Syntax
Description
send
Output a single measurement message.
send [aaa]
Output a single measurement message from
the device with the defined address:
aaa
= address of target device, range
0...255.
Example:
send 5
T= 22.8 'C RH= 39.8 %RH Td=
8.4 'C Tw= 14.6 'C h=
40.5 kJ/kg
Table 10 Send Command
Syntax
Description
r
Start the continuous outputting of
measurement values as an ASCII text string
to the serial line.
Example (measurement message in default format):
r
T= 22.8 'C RH= 39.5 %RH Td=
8.3 'C Tw= 14.5 'C h=
40.4 kJ/kg
T= 22.8 'C RH= 39.5 %RH Td=
8.3 'C Tw= 14.5 'C h=
40.4 kJ/kg
...
Table 11 R Command
28
4 Service Port