Agilent Technologies N8201A User Manual
Page 53
![background image](/manuals/40830/53/background.png)
Hardware Setup and Configuration
2
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz
49
Measurement Procedure
4
Tune the Spectrum Analyzer to the following:
•
Frequency: 7.5 MHz
•
Amplitude: 10 dBm
•
Span: 100 kHz
5
Tune the downconverter to the following:
•
Center Frequency: 3.5 GHz
•
Attenuator: 20 dBm
•
RF Enabled: Enabled
•
Source: Internal
6
Tune the Source to the following:
•
Frequency: 3.5 GHz
•
Amplitude: 0 dBm
•
RF On
Verify that a 7.5 MHz signal is present on the spectrum analyzer.