Agilent Technologies N8201A User Manual
Page 50
46
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz
2
Hardware Setup and Configuration
Measurement Procedure
5
Tune the Spectrum Analyzer to the following:
•
Frequency: 7.5 MHz
•
Amplitude: 10 dBm
•
Span: 100 kHz
6
Tune the downconverter to the following:
•
Level: Attenuator: 20 dBm
•
RF Enabled: Enabled
•
Input RF: Signal Source: 50 MHz Cal
•
Frequency: 50 MHz
7
Verify that a 7.5 MHz signal is present on the spectrum analyzer.