Preventive Maintenance
5
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz
99
Half-Rack-Width Instrument
Bench Top Instrument
To remove a half-width instrument from a system rack
1 Power off the system.
•
For details see the system installation
guide.
2 Remove the selected instrument’s power cord
from the power strip in the rack.
3 The instrument is attached to the half-rack width
instrument beside it; remove that instrument’s
power cord from the power strip also.
•
The instruments are secured together
by lock links at the front and rear. The
lock links at the rear attach with screws.
The lock links at the front hook together.
4 Remove the power cord and other cables from the
front and rear of both instruments.
•
Note the location of cables for
re-installation.
5 Remove the four corner screws on the front of the
rack panel that secures the instruments in place.
•
The screws are located near the corners
of the face of the instrument.
•
Use a #2 Phillips screwdriver.
6 Slide both instruments, as a single unit, out from
the front of the rack and set them on a secure, flat
surface.
To remove an instrument from a bench top system
1 Power off each instrument in the system.
2 Unplug the selected instrument’s power
cord from the AC power supply.
3 Remove the power cord and other cables
from the front and rear of the instrument.
•
Note the location of cables for re-installation.