15. Specifications
MS-70
MX-50
MF-50
ML-50
Measurement method
400 W halogen lamp, thermogravimetric analysis
Drying temperature
range at sample pan
( Increments)
30°C to 200°C
(1°C )
50°C to 200°C
(1°C )
Heating pattern
Standard drying, Ramp drying,
Step drying, Quick drying
Standard drying,
Quick drying
Temperature calibration
By Accessory AX-MX-43
-
Sample weight range
0.1 g to 71 g
0.1 g to 51 g
Accuracy: Repeatability of measurement, (Standard deviation)
Moisture
content 1
over 5 g sample
0.01 %
0.02 %
0.05 %
0.1 %
over 1 g sample
0.05 %
0.1 %
0.2 %
0.5 %
Weighing mode
0.0005 g
0.001 g
0.002 g
0.005 g
Minimum reading
Moisture
content
0.001
%,
0.01
%,
0.1
%
0.01
%,
0.1
%
0.05%,
0.1
%,
1
%
0.1%,
1
%
Weighing
mode
0.0001
g
0.001g
0.002g
0.002g
Measurement programs
Sample weight and termination value is automatically set with
ACCURACY and % display. When drying rate reaches the
termination value, measurement is automatically completed. ( 2)
Standard mode
Standard drying, Ramp drying, Step drying
Standard drying
Sample weight and termination value is automatically set with
ACCURACY and % display. When drying rate reaches the
termination value, measurement is automatically completed. ( 2)
Quick mode
Quick drying
When drying rate is less than preset termination value,
measurement is automatically completed. ( 2)
Automatic mode
Standard drying, Ramp drying, Step drying
Standard drying
After heating for the preset drying time, measurement is
automatically stopped. (1min. to 480 min.)
Timer mode
Standard drying, Ramp drying, Step drying
Standard drying
When pressing the key at any time, measurement is stopped and
the result is decided.
Analys
is mode
Manual mode
Standard drying, Ramp drying, Step drying
Standard drying
Moisture content (Wet-base)
Moisture content (Dry-base, Atro)
Dry
content
Ratio
Measurement unit
Weight (g)
Number of memory
20 sets
10 sets
5 sets
Data memory function
100 results
50 results
30 results
Communication function
RS-232C serial interface
Operation environment
5°C to 40°C (41°F to 104°F), 85%RH or less (no condensation)
Sample pan
φ85 mm
1: After preheating the analyzer, the data can be obtained with approximately 5 g test
sample (Sodium tartrate dihydrate) in standard mode (MID.), standard drying, 160 °C
2: When change of moisture content per one minute reaches the preset termination
value, the measurement is completed.