Specifications – A&D ML-50 User Manual

Page 70

background image

68

MS-70 / MX-50 / MF-50 / ML-50

15. Specifications

MS-70

MX-50

MF-50

ML-50

Measurement method

400 W halogen lamp, thermogravimetric analysis

Drying temperature

range at sample pan

( Increments)

30°C to 200°C

(1°C )

50°C to 200°C

(1°C )

Heating pattern

Standard drying, Ramp drying,

Step drying, Quick drying

Standard drying,

Quick drying

Temperature calibration

By Accessory AX-MX-43

-

Sample weight range

0.1 g to 71 g

0.1 g to 51 g

Accuracy: Repeatability of measurement, (Standard deviation)
Moisture

content 1

over 5 g sample

0.01 %

0.02 %

0.05 %

0.1 %

over 1 g sample

0.05 %

0.1 %

0.2 %

0.5 %

Weighing mode

0.0005 g

0.001 g

0.002 g

0.005 g

Minimum reading

Moisture

content

0.001

%,

0.01

%,

0.1

%

0.01

%,

0.1

%

0.05%,

0.1

%,

1

%

0.1%,

1

%

Weighing

mode

0.0001

g

0.001g

0.002g

0.002g

Measurement programs

Sample weight and termination value is automatically set with

ACCURACY and % display. When drying rate reaches the

termination value, measurement is automatically completed. ( 2)

Standard mode

Standard drying, Ramp drying, Step drying

Standard drying

Sample weight and termination value is automatically set with

ACCURACY and % display. When drying rate reaches the

termination value, measurement is automatically completed. ( 2)

Quick mode

Quick drying

When drying rate is less than preset termination value,

measurement is automatically completed. ( 2)

Automatic mode

Standard drying, Ramp drying, Step drying

Standard drying

After heating for the preset drying time, measurement is

automatically stopped. (1min. to 480 min.)

Timer mode

Standard drying, Ramp drying, Step drying

Standard drying

When pressing the key at any time, measurement is stopped and

the result is decided.

Analys

is mode

Manual mode

Standard drying, Ramp drying, Step drying

Standard drying

Moisture content (Wet-base)

Moisture content (Dry-base, Atro)

Dry

content

Ratio

Measurement unit

Weight (g)

Number of memory

20 sets

10 sets

5 sets

Data memory function

100 results

50 results

30 results

Communication function

RS-232C serial interface

Operation environment

5°C to 40°C (41°F to 104°F), 85%RH or less (no condensation)

Sample pan

φ85 mm

1: After preheating the analyzer, the data can be obtained with approximately 5 g test

sample (Sodium tartrate dihydrate) in standard mode (MID.), standard drying, 160 °C

2: When change of moisture content per one minute reaches the preset termination

value, the measurement is completed.