A&D ML-50 User Manual

Page 60

background image

58

MS-70 / MX-50 / MF-50 / ML-50

Calibration Report Example for the Weighing Sensor Adapted to GLP, GMP and ISO


Preset the following parameters to print data to AD-8121B

Device Parameter

Description

Interval

pU5e 1 1 Approx. two seconds interval in each line.

Analyzer

Output format
adapted to GLP,
GMP and ISO

info 1
or
info 2

Calibration report is printed at "Calibration of the
Weighting Sensor" and "Calibration of Drying
Temperature".

AD-8121B

MODE 3

Dump print (Received data is printed as it is)

1: Factory settings



...........

A & D .........

Manufacture

MODEL

......

MX-50 .........

Model

S/N

.....

K1234567 .........

Serial number

ID

.......

LAB-123 .........

ID number

DATE

..

2004/09/30 .........

Date

TIME

....

13:57:24 .........

Time

CALIBRATED

..........

WEIGHT

CAL.WEIGHT

.......

20.000

..

g

SIGNATURE

.........

Signature

- - - - - - - -






}

Calibration type

}

Calibration mass