A&D ML-50 User Manual
Page 60

58
MS-70 / MX-50 / MF-50 / ML-50
Calibration Report Example for the Weighing Sensor Adapted to GLP, GMP and ISO
 
Preset the following parameters to print data to AD-8121B 
Device Parameter
Description
Interval
pU5e 1 1 Approx. two seconds interval in each line.
Analyzer
Output format 
adapted to GLP, 
GMP and ISO 
info 1 
or 
info 2 
Calibration report is printed at "Calibration of the 
Weighting Sensor" and "Calibration of Drying 
Temperature". 
AD-8121B
MODE 3
Dump print (Received data is printed as it is)
1: Factory settings
 
 
 
...........
A & D .........
Manufacture
MODEL
......
MX-50 .........
Model
S/N
.....
K1234567 .........
Serial number
ID
.......
LAB-123 .........
ID number
DATE
..
2004/09/30 .........
Date
TIME
....
13:57:24 .........
Time
CALIBRATED
..........
WEIGHT
CAL.WEIGHT
.......
20.000
..
g
SIGNATURE
.........
Signature
- - - - - - - -
 
 
 
 
 
 
}
Calibration type
}
Calibration mass