Function table – A&D ML-50 User Manual
Page 63

MS-70 / MX-50 / MF-50 / ML-50
61
13. Function Table
The function table can store the following parameters to control the analyzer.
Details of the Function Table
Item and Display Symbol
Param-
eter
Description
Clock Cl
adj
Set date and time for the built-in clock.
Refer to "5.2. Setting the clock and calendar"
0
1 Dot " . "
Decimal point dp
1
Comma " , "
Select decimal point of data.
0
1 Key mode
Data is output by the
ENTER
key
1
Auto print mode Data is output after measurement.
Data output
mode
prt
2
Stream mode
Data is output continuously during
measurement.
0
1 Not used.
Data memory
function
data
1
Data is stored at each measurement
0
1 Moisture content is output.
Form selection 5-d
1
Moisture content and temperature are output. 2
0
Continuous output
Interval
pU5e
1
1 Output with approx. two seconds interval in each line.
0
1 Not used
1
To output (print) items with "analyzer
information", "measurement program",
"measurement data" and ""signature
space" at one time. 3
Output format
adapted to
GLP, GMP and
ISO
info
2
To output (print) items specified from
"analyzer information", "measurement
program" and "signature space". 4
Output of
calibration
report. 5
ID number
id
Set ID number. Used for the calibration report.
Factory
settings
Clr
Resets the analyzer to the factory settings.
1: Factory settings
2: Use a computer because the AD-8121B printer can not print this correctly.
RsTemp and RsFig contained in the software "WinCT-Moisture" that is the standard
accessory of the MS-70 and MX-50 can not output this correctly.
3: Refer to "9.1.1. Example To Print The Whole Data At One Time".
4: Refer to "9.1.2. Example To Print Selected Items".
5: Calibration report is outputted in "Calibration of the Weighting Sensor" and
"Calibration of Drying Temperature".