A&D ML-50 User Manual

Page 41

background image

MS-70 / MX-50 / MF-50 / ML-50

39

Set the Minimum Scale Value of the % Display

13. Select the % display with the

SELECT

key.

Select 0.01 [%] with the

or

key.


Set the Minimum Scale Value of the Gram Display

14. Select the gram display with the

SELECT

key.

Select 0.001 [g] with the

or

key.



Store the Parameters and Finish the Operation

15. Press the

ENTER

key to store the new parameters of

the measurement program to program number 3.
Pressing the key, the weighing mode is automatically
displayed.
When PROG 3 is recalled, the settings can be used.


To cancel new parameters and return to weighing

mode, press the

RESET

key.




Timer Mode

Refer to page 37 for Standard Mode, Automatic Mode or Manual Mode


This explanation uses the following parameters and sample displays of MX-50.






"Temperature 1" can be set higher than "temperature 2" in step drying.

Common

Items

Program

number ..........................................6 ( PROG 6 )

Analysis

mode ..............................................Timer mode (Symbol: U-t )

Drying

program

Heating pattern....................Step drying (

)

Drying temperature 1 ..........160°C

Drying temperature 2 ..........120°C

Time 1..................................5.0 minutes

Time 2..................................10.0 minutes

Measurement

unit ..............................................Moisture content

Minimum scale value during measurement ..0.01 %

Minimum scale value of the gram display.....0.001 g

Sample

quantity............................................Approximately 5 g

Data

memory

function ..................................Not used


Time 1

Drying temperature 2

Drying temperature 1

Stage 1

Stage 2

Time 2

% MOIST

D

/W

/D

0.000

g

160°C

Gram display of
weighing mode

"0."0"1

%

Minimum scale
value of % display

"0."0"0"1

g

Minimum scale
value of gram
display