MS-70 / MX-50 / MF-50 / ML-50
29
How to Select quick mode (Extracts)
Press
the
SELECT
key during the gram display.
Display
quc
using the
↓
or
↑
key.
ML-50 is does not display heating pattern.
How to Select a Heating pattern without quick mode (Extracts)
Press
the
SELECT
key to select temperature
during the gram display.
Press
the
PROGRAM
key to select a symbol of
heating pattern are displayed in order.
ML-50 can select standard drying and quick drying only.
Display examples during settings
Drying Temperature at the Sample Pan
Drying temperature range for MS-70....................................... 30°C to 200°C, 1°C interval.
Drying temperature range for MX-50, MF-50 and ML-50 ....... 50°C to 200°C, 1°C interval.
When a measurement is started and one hour passes, the maximum temperature is
automatically regulated to 160°C for safety.
7.1.5. Measurement
Unit
Unit Formula
Display
Moisture content is based on
wet sample mass
*
1
W - D
W
% MOIST
D
/W
/D
Moisture content (Atro) is based on
dried sample mass
W - D
D
% MOIST
D
/D
Dry content
D
W
% RATIO
D/W
/D
Ratio
*
2
W
D
% RATIO
W/D
Gram value
-
g
W: Wet sample mass
D: Dried sample mass
*
1: Factory settings
*
2: When result reaches to 999%, measurement is stopped.
x 100
x 100
x 100
x 100
Heating pattern
Time
Time2
Time1
Temperature
Temperature2
Temperature
Heating pattern
Heating pattern
Temperature
"q"u"c
Quick mode
Heating pattern
"1"0"5°C