Analyzing mode of the automatic mode, Analyzing mode of the timer mode – A&D ML-50 User Manual
Page 30

28
MS-70 / MX-50 / MF-50 / ML-50
7.1.2. Analyzing mode of the Automatic Mode
When the change of moisture content per one minute is less than the preset value,
the measurement is automatically completed.
Range
Preset Termination
value to complete
measurement
MS-70 MX-50 MF-50 ML-50
2.00
%/min
1.00
%/min
0.50
%/min
Available Available
0.20
%/min
Available
range
range
0.10
%/min
Available
range
(Factory setting)
(Factory setting)
0.05
%/min
range
(Factory setting)
0.02
%/min
(Factory setting)
0.01
%/min
0.005
%/min
0.002
%/min
0.001
%/min
Unavailable item
7.1.3. Analyzing mode of the Timer Mode
Sample is dried for a preset analysis time.
Drying Time
1 minute to 480 minutes.
Setting interval :
1 minute during 1 minute to 60 minutes.
5 minutes during 60 minutes to 480 minutes.
Factory setting:
10 minutes.
7.1.4. Drying Program (Heating Pattern and Drying Temperature)
Heating pattern
Standard Mode,
Automatic Mode,
Manual Mode
Timer Mode
Quick Mode
Standard drying
-
Ramp drying
-
Step drying
-
Quick drying
-
-
"Temperature 1" can be set higher than "temperature 2" in step drying.
ML-50, can select standard drying and quick drying only, does not display heating pattern.
Temperature
Temperature
Time
Temperature
Temperature
Time1
Time2
Time
Stage 1
Temperature1
Time
Temperature2
Stage 2
Temperature1
Time1
Temperature2
Stage 1
Stage 2
Time2
Temperature
200°C
3 min.