beautypg.com

Procedures to store a measurement program, Standard drying – A&D ML-50 User Manual

Page 32

background image

30

MS-70 / MX-50 / MF-50 / ML-50

7.2. Procedures to Store a Measurement Program

7.2.1. Standard

Drying

This explanation uses the following parameters and sample displays of MX-50.






Common

Items

Program

number ..........................................2 ( PROG 2 )

Drying

program

Heating

pattern....................Standard drying (

)

Drying temperature .............160°C

Measurement

unit ..............................................Moisture content

Minimum scale value during measurement ..0.01 %

Minimum scale value of gram display...........0.001 g

Sample

quantity............................................Approximately 5 g

Data

memory

function ..................................Not used

Items for Standard Mode

Analysis

mode ..............................................Standard mode (Symbol: 5td )

ACCURACY .................................................MID.

Analyzing mode to finish a measurement.........Automatic setting by ACCURACY

Items

for

Automatic

Mode

Analysis

mode ..............................................Automatic mode (Symbol: U-a )

Analyzing mode to finish measurement............0.05 %/min

Items

for

Timer

Mode

Analysis

mode ..............................................Timer mode (Symbol: U-t )

Analyzing mode to finish measurement............10 minutes

Items for Manual Mode

Analysis mode ..............................................Manual mode (Symbol: U-m )



1. Display the gram unit (of the weighing mode).

Select a Program Number to Edit the Measurement Program

2. Press

the

PROGRAM

key and press the

or

key to select a program number.

3. Press

the

ENTER

key to use the number.

4. The analyzer displays

end

and returns to the

weighing mode.

Caution If the data memory function is active, the data

memory number (MEM) is displayed in place of the
program number (PROG).

Drying temperature

Time

Drying temperature

Standard Mode, Automatic Mode, Manual Mode

Timer Mode

% MOIST

D

/W

/D

PROG 1

0.000

g

PROG "2

end

PROG 2

0.000

g

Program number

This manual is related to the following products: