beautypg.com

A&D ML-50 User Manual

Page 29

background image

MS-70 / MX-50 / MF-50 / ML-50

27

7.1.1. ACCURACY of the Standard Mode and Quick Mode

ACCURACY of measurement can be set either

HI

,

MID.

or

LO

.

The sample quantity is automatically selected by ACCURACY.

The termination value of the analyzing mode is automatically selected by ACCURACY

and minimum scale value of % display.

The analyzing mode is the program to finish the measurement when a change of

moisture content per one minute becomes smaller than a preset termination value.

The settings are as follows: Specify an ACCURACY.

Drying rate: Change of moisture content per one minute [%/min]


Standard Mode

ACCURACY

Model

Minimum

scale

HI

MID.

LO

0.001 %

0.01 %/min 0.02 %/min 0.05 %/min

0.01 %

0.02 %/min 0.05 %/min 0.10 %/min

MS-70

0.1 %

0.10 %/min 0.20 %/min 0.50 %/min

0.01 %

0.02 %/min 0.05 %/min 0.10 %/min

MX-50

0.1 %

0.10 %/min 0.20 %/min

0.05 %

0.05 %/min 0.10 %/min

0.1

%

MF-50

1

%

0.1

%

Preset
Termination
value

ML-50

1

%

0.10 %/min 0.20 %/min

0.50 %/min

Sample quantity

10 g

5 g

1 g

Use Precise

result

Quick

measurement


Quick Mode

ACCURACY

Model

Minimum

scale

HI

MID.

LO

0.001 %

0.02 %/min 0.05 %/min 0.05 %/min

0.01 %

0.05 %/min 0.10 %/min 0.20 %/min

MS-70

0.1 %

0.10 %/min 0.20 %/min 0.50 %/min

0.01 %

0.05 %/min 0.10 %/min 0.20 %/min

MX-50

0.1 %

0.10 %/min 0.20 %/min 0.50 %/min

0.05

%

0.1

%

MF-50

1

%

0.10 %/min 0.20 %/min 0.50 %/min

0.1

%

Preset
Termination
value

ML-50

1

%

0.20 %/min 0.50 %/min 1.00 %/min

Sample quantity

5 g

2 g

1 g

Use Precise

result

Quick

measurement

This manual is related to the following products: