MS-70 / MX-50 / MF-50 / ML-50

Select an Analysis Mode

5. Press

the

SELECT

key to select a symbol, it

will blink. Select

U-t

of the timer mode with

the

or

key. (Press these keys to select it)


Set the Heating Pattern

6. Select

drying

temperature with the

SELECT

key.

Select ramp drying

of the heating pattern

with the

PROGRAM

key.


Set the Final Drying Temperature

7. Set 160°C with the

or

key.



Set the Ramp Time

8. Set 5.0 minutes with the

or

key.



Set the Analysis Time

9. Set 10.0 minutes with the

or

key.



Set a Measurement Unit

10. Select a measurement unit with the

SELECT

key. Select moisture content (based on a wet
sample ) with the

or

key.


Set the Minimum Scale Value of the % Display

11. Select the % display with the

SELECT

key.

Select 0.01 [%] with the

or

key.


Set the Minimum Scale Value of the Gram Display

12. Select gram display with the

SELECT

key.

Select 0.001 [g] with the

or

key.


Store the Parameters and Finish the Operation

13. Press the

ENTER

key to store the new parameters

of the measurement program to program number 4.
Pressing the key, the weighing mode is
automatically displayed. When PROG 4 is recalled,
the settings can be used.

To cancel the new parameters and return to the

weighing mode, press the

RESET

key.

"U"-"t

0.000

g

Analysis mode

1

0.000

g

"1"0"5°C

Ramp Time

"5."0 min

2

0

"%

"M"O"I"S"T
"/"W

Measurement unit

"0."0"1

%

Minimum scale
value of % display

"0."0"0"1

g

Minimum scale
value of gram
display

0.000

g

160°C

10.0 min

"1"0."0 min

Analysis time

0.000

g

"1"6"0°C

Final drying temperature

Ramp drying

Gram display of
weighing mode