A&D ML-50 User Manual
Page 33

MS-70 / MX-50 / MF-50 / ML-50
31
Select an Analysis Mode
5. Press
the
SELECT
key to select a symbol, it will
blink. Select an analysis mode using the
↓
or
↑
key.
If standard mode is used, select
5td
.
If automatic mode is used, select
U-a
.
If timer mode is used, select
U-t
.
If manual mode is used, select
U-m
.
Set ACCURACY for the Standard Mode
6. Select ACCURACY with the
SELECT
key.
Select MID. with the
↓
or
↑
key.
When
ACCURACY
is selected, HI, MID. or LO blinks.
Set the Heating Pattern
7. Select
drying
temperature with the
SELECT
key.
Select standard drying
of the heating
pattern with the
PROGRAM
key.
Set the Drying Temperature
8. Set 160°C with the
↓
or
↑
key.
Set the Preset Termination Value for Automatic Mode
9. Select the preset termination value with the
SELECT
key.
Select 0.05 [%/min] with the
↓
or
↑
key.
Set the Analysis Time for the Timer Mode
10. Select time with the
SELECT
key.
Select 10.0 [min] with the
↓
or
↑
key.
Set the Measurement Unit
11. Select a measurement unit with the
SELECT
key. Select a moisture content (based on a wet
sample ) with the
↓
or
↑
key.
Set the Minimum Scale Value of the % Display
12. Select % display with the
SELECT
key.
Select 0.01 [%] with the
↓
or
↑
key.
0.000
g
"1"0."0 min
Analysis Time
ACCURACY
MID.
ACCURACY
"0."0"5
%/min
Preset termination value
Standard
mode
Automatic
mode
Timer mode
Standard mode
Manual mode
"U"-"a
0.000
g
Analysis mode
0.000
g
"1"6"0°C
0
"%
"M"O"I"S"T
"/"W
Measurement unit
"0".0"1
%
Minimum scale
value of %
display
0.000
g
Standard drying
Drying temperature
Automatic mode
Timer mode
Manual mode
"1"0"5°C