beautypg.com

A&D ML-50 User Manual

Page 33

background image

MS-70 / MX-50 / MF-50 / ML-50

31

Select an Analysis Mode

5. Press

the

SELECT

key to select a symbol, it will

blink. Select an analysis mode using the

or

key.

If standard mode is used, select

5td

.

If automatic mode is used, select

U-a

.

If timer mode is used, select

U-t

.

If manual mode is used, select

U-m

.


Set ACCURACY for the Standard Mode

6. Select ACCURACY with the

SELECT

key.

Select MID. with the

or

key.

When

ACCURACY

is selected, HI, MID. or LO blinks.

Set the Heating Pattern

7. Select

drying

temperature with the

SELECT

key.

Select standard drying

of the heating

pattern with the

PROGRAM

key.


Set the Drying Temperature

8. Set 160°C with the

or

key.



Set the Preset Termination Value for Automatic Mode

9. Select the preset termination value with the

SELECT

key.

Select 0.05 [%/min] with the

or

key.



Set the Analysis Time for the Timer Mode

10. Select time with the

SELECT

key.

Select 10.0 [min] with the

or

key.



Set the Measurement Unit

11. Select a measurement unit with the

SELECT

key. Select a moisture content (based on a wet
sample ) with the

or

key.


Set the Minimum Scale Value of the % Display

12. Select % display with the

SELECT

key.

Select 0.01 [%] with the

or

key.


0.000

g


"1"0."0 min

Analysis Time

ACCURACY

MID.

ACCURACY

"0."0"5

%/min

Preset termination value

Standard

mode

Automatic

mode

Timer mode

Standard mode
Manual mode

"U"-"a

0.000

g

Analysis mode

0.000

g

"1"6"0°C

0

"%

"M"O"I"S"T
"/"W

Measurement unit

"0".0"1

%

Minimum scale

value of %

display

0.000

g

Standard drying

Drying temperature

Automatic mode
Timer mode
Manual mode

"1"0"5°C

This manual is related to the following products: