beautypg.com

Program number, Storing a measurement program to a program number – A&D ML-50 User Manual

Page 27

background image

MS-70 / MX-50 / MF-50 / ML-50

25

6.3. Program Number

The measurement conditions of all program numbers are set to the standard mode at the

factory. The analyzer can store and recall proper individual settings for each sample with
the program number (PROG No.).

MS-70 / MX-50

20 sets

PROG 1 to 20

MF-50

10 sets

PROG 1 to 10

ML-50

0

5 sets

PROG 1 to 5

The same measurement program is stored in all program

numbers with factory settings.

Analysis

mode..................... Standard mode

Heating

pattern ................... Standard drying

Caution If the data memory function is active, the data memory number (MEM) is

displayed, in place of the program number (PROG).

6.3.1. Storing a Measurement Program to a Program Number

Displaying or recalling a PROG number, a measurement program can be renewed.

1. Press and hold the

PROGRAM

key in gram display.

2. Press

the

or

key to select a program number


3. Press

the

ENTER

key to use the selected number.



4. Press

the

SELECT

key to edit the parameters.

5. Edit parameters of a measurement program.

Refer to "7.Selection of Measurement Method"

6. Press the following key to return to the gram display.

ENTER

key......Stores parameters to the selected number.

RESET

key......Cancels the operation and returns to the

weighing mode.

6.3.2. Recalling a Measurement Program with a Program Number

Stored

measurement

programs can be recalled with a program number.

1. Press and hold the

PROGRAM

key in the gram display.


2. Press

the

or

key to select a program number


3. Press

the

ENTER

key to use the selected number.



Program number

Memory number

PROG "1

0.000

g

PROG "2

end

PROG 2

0.000

g

Program number

Edit measurement

program.

end

PROG 2

0.000

g

PROG "1

0.000

g

PROG "2

end

PROG "2

0.000

g

Program number

This manual is related to the following products: