A&D ML-50 User Manual

Page 36

background image

34

MS-70 / MX-50 / MF-50 / ML-50

Select an Analysis Mode

5. Press

the

SELECT

key to select a symbol, it will

blink. Select an analysis mode using the

or

key.

If standard mode is used, select

5td

.

If automatic mode is used, select

U-a

.

If manual mode is used, select

U-m

.


Set ACCURACY for Standard Mode

6. Select ACCURACY with the

SELECT

key.

Select MID. with the

or

key.

When

ACCURACY

is selected, HI, MID. or LO blinks.

Set the Heating Pattern

7. Select

drying

temperature with the

SELECT

key.

Select ramp drying

of heating pattern with

the

PROGRAM

key.


Set the Final Drying Temperature

8. Set 160°C with the

or

key.



Set the Ramp Time

9. Set 5.0 minutes with the

or

key.



Set the Preset Termination Value for Automatic Mode

10. Select preset termination value with the

SELECT

key. Select 0.05 [%/min] with the

or

key.


Set the Measurement Unit

11. Select measurement unit with the

SELECT

key.

Select a moisture content (based on a wet
sample ) with the

or

key.


Set the Minimum Scale Value of the % Display

12. Select % display with the

SELECT

key.

Select 0.01 [%] with the

or

key.


Set the Minimum Scale Value of the Gram Display

13. Select gram display with the

SELECT

key.

Select 0.001 [g] with the

or

key.


Automatic

mode

0

"%

"M"O"I"S"T
"/"W

"0."0"1

%

"0."0"0"1

g

0.000

g

"1"6"0°C

ACCURACY

MID.

"0."0"5

%/min

0.000

g

5.0 min

0.000

g

""1"0"5°C

Measurement unit

Minimum scale
value of % display

Minimum scale
value of gram
display

Final drying temperature

Ramp time

Ramp drying

ACCURACY

Preset termination value

Standard mode
Manual mode

"U"-"a

0.000

g

Standard

mode

Automatic mode
Manual mode

Analysis mode