Boonton 9240 Series RF Voltmeter

Table 5-1. SCPI COMMAND SUMMARY (continued)


CALCulate:MATH:ARGB

Set or return the second argument to be used for channel math
operations.


CALCulate:MATH:DATA?

Returns the channel math result.


CALCulate:MATH:OPERator

Set or return the channel math operator.


CALCulate:MODE Set/return

remote measurement mode

= NORMal, FAST, FILTered


CALCulate[1|2]:REFerence:COLLect Set the current reading to be the ratiometric measurement reference

level.


CALCulate[1|2]:REFerence:DATA

Set or return the ratiometric measurement reference level in dBm.

= -99.99 to +99.99 dBm


CALCulate[1|2]:REFerence:STATe

Set or return the ratiometric measurement mode state.

= 0, 1, OFF, ON


CALCulate[1|2]:STATe Enable

currently selected channel allowing measurements to be

made.

= 0, 1, OFF, ON


CALCulate[1|2]:UNITs

Change channel units. = DBM, Watts, Volts,
DBV, DBMV, DBUV


CALibration[1|2]:ZERO Start

zero

process.


DISPlay:ACTive

Set/return the active channel for talk commands.


DISPlay:CLEar

Clear measurement data and display.


DISPlay:LIN:RESolution

Set number of significant digits for linear displays and remote
return values. = 3 to 5


DISPlay:LOG:RESolution Set

number of decimal places for log displays and remote return

values. = 1 to 3


DISPlay:LABel:MODE

Enables/disables the user message display.
= 0, 1, OFF, ON


DISPlay:Label:TEXTA

Displays a text message in the first label field.
= A to Z, a to z, 0 to 9 ! @ # $ % ^ & *

( ) _ - + = { } [ ] ? / < > : .


DISPlay:Label:TEXTB

Displays a text message in the second label field.
= A to Z, a to z, 0 to 9 ! @ # $ % ^ & *

( ) _ - + = { } [ ] ? / < > : .


DISPlay:Label:TEXTC

Displays a text message in the third label field.
= A to Z, a to z, 0 to 9 ! @ # $ % ^ & *

( ) _ - + = { } [ ] ? / < > : .


Remote Operation