6 specifications, Voltage probe inputs, Features – Boonton 9240 RF Voltmeter User Manual
Page 17: 6 specifications -7, Voltage probe inputs -7, Features -7, Boonton 9240 series rf voltmeter, General information 1-7

Boonton 9240 Series RF Voltmeter
1.6 Specifications
Performance specifications for the 9240 Series are listed in Table 1-2.
Table 1-2 9240 Series Performance Specifications
(Specifications are subject to change without notice)
VOLTAGE PROBE INPUTS
Frequency Range:
10 Hz to 1.2 GHz
1
(
1
Voltage Probe dependant )
Voltage Range:
200 µV to 10V
Dynamic Range: 90 dB
Waveform response:
RMS to 30mV, calibrated in RMS of a sinewave above 30mV
(RMS to 10V and 700MHz with 100:1 divider)
Crest Factor:
Direct Input: Level 300uV 1mV 3mV 10mV 30mV
CF 140 42 14 4.2 1.4
With Divider: Level 30mV 100mV 300mV 1V 3V
CF 140 42 14 4.2 1.4
Input Impedance:
Refer to Figure 1-2
Maximum AC Input:
10V
Maximum DC Input:
200V
FEATURES
Display:
Menu-driven 20 character x 4 line LCD
Display Units:
VOLTS, dBV, dBmV, dBµV, dBnV, WATTS, dBW, dBm
2
, dBµW,
dBnW, dBr, %
(
2
0dBm = 1mW calculated from V
2
/Z
o
where Z
o
is the selected reference impedance)
Z
o
Impedance:
Any selected reference value from 5 to 2500
Ω
Peak Display:
1.414 times Voltage display or 3.01 dB added to dB display
Display Resolution:
0.001 dB or 5 digits (in Volts mode)
Display Offset:
-99.99 dB to +99.99 dB in 0.01 dB steps
Limiting:
Individual high and low limit thresholds, -99.99 dB to +99.99 dB
Ranging:
Manual (8 ranges) or autoranging
Filtering:
Filter times to 20.00 seconds in 0.05 second increments
Zeroing:
Automatic function; calculates, stores, and applies zero corrections to
each
range
Reference Level:
-99.99 dB to +99.99 dB in 0.01 dB steps for dBr measurements
may be express in % in linear mode.
Recorder Ouput:
For linear voltage readout, 10V full scale proportional to indicated
voltage over each decade. Ranging occurs at 10% of full scale. For
decibel readouts, scale factor is 1V per 10dB change over the entire
dynamic
range.
General Information
1-7