Boonton 9240 Series RF Voltmeter

5.5.5 FETCh Queries

The FETCh? group of queries is used to return specific measurement data from a measurement cycle that has been INITiated
and is complete or free-running. FETCh? performs the data output portion of the measurement. FETCh? does not start a
new measurement, so a series of FETCh? queries may be used to return more than one set of processed measurements from a
complete set of acquired data. FETCh? usually returns the current value of measurements, and should be used anytime free
running data acquisition is taking place (INITiate:CONTinuous ON). If FETCh? is used for single measurements
(INITiate:CONTinuous OFF), no data will be returned until a measurement has been INITiated and is complete.


FETCh:CW:VOLTage?

Description:

Return current average amplitude reading in channel units.

Syntax: FETCh[1|2]:CW:VOLTage?

Returns:

CC, average voltage

Where CC is the measurement condition code.


FETCh:KEY?

Description:

Return the key code of the last key depressed; e.g. MENU = 1.

Syntax:

FETCh:KEY?

Returns: key

code

Key

Code

Menu

1

Sensor

2

FREQ

4

AVG

8

Zero/Cal

16

REF

Level

32

Up

Arrow

64

Left

Arrow

128

Enter

256

Right

Arrow

512

Down

Arrow

1024






Remote Operation