5 fetch queries, Fetch:cw:voltage, Fetch:key – Boonton 9240 RF Voltmeter User Manual
Page 67: 5 fetch queries -17, Fetch:cw:voltage? -17, Fetch:key? -17, Boonton 9240 series rf voltmeter, Remote operation 5-17

Boonton 9240 Series RF Voltmeter
5.5.5 FETCh Queries
The FETCh? group of queries is used to return specific measurement data from a measurement cycle that has been INITiated
and is complete or free-running. FETCh? performs the data output portion of the measurement. FETCh? does not start a
new measurement, so a series of FETCh? queries may be used to return more than one set of processed measurements from a
complete set of acquired data. FETCh? usually returns the current value of measurements, and should be used anytime free
running data acquisition is taking place (INITiate:CONTinuous ON). If FETCh? is used for single measurements
(INITiate:CONTinuous OFF), no data will be returned until a measurement has been INITiated and is complete.
FETCh:CW:VOLTage?
Description:
Return current average amplitude reading in channel units.
Syntax: FETCh[1|2]:CW:VOLTage?
Returns:
CC, average voltage
Where CC is the measurement condition code.
FETCh:KEY?
Description:
Return the key code of the last key depressed; e.g. MENU = 1.
Syntax:
FETCh:KEY?
Returns: key
code
Key
Code
Menu
1
Sensor
2
FREQ
4
AVG
8
Zero/Cal
16
REF
Level
32
Up
Arrow
64
Left
Arrow
128
Enter
256
Right
Arrow
512
Down
Arrow
1024
Remote Operation
5-17