Calculate:math:arga, Calculate:math:argb, Calculate:math:data – Boonton 9240 RF Voltmeter User Manual
Page 62: Calculate:math:operator, Calculate:mode, Calculate:math:arga -12, Calculate:math:argb -12, Calculate:math:data? -12, Calculate:math:operator -12, Calculate:mode -12

Boonton 9240 Series RF Voltmeter
CALCulate:MATH:ARGA
Description:
Set or return the first argument to be used for channel math operations.
Syntax:
CALCulate:MATH:ARGA
Argument:
CALCulate:MATH:ARGB
Description:
Set or return the second argument to be used for channel math operations.
Syntax:
CALCulate:MATH:ARGB
Argument:
CALCulate:MATH:DATA?
Description:
Returns the channel math result.
Syntax:
CALCulate:MATH:DATA?
CALCulate:MATH:OPERator
Description:
Set or return the channel math operator.
Syntax:
CALCulate:MATH:OPERator
Argument:
CALCulate:MODe
Description:
Set or return the system remote measurement mode. These measurement modes are only available
under bus control. NORMal is the default mode of operation at approximately 20 readings per
second. FAST returns data at over 200 readings per second. FILTered will only return data when
the filter is full. FILTered works best when INITiate:CONTinuous is OFF and a trigger is issued.
In this case the measurement will not be returned until the filter is full. When the instrument is
returned to local the MODE defaults to NORMal.
Syntax:
CALCulate:MODe
Argument:
Remote Operation
5-12