5.5.13 SYSTem Subsystem

The SYSTem group of commands is used to control system-level functions not directly related to instrument measurement
performance. SYSTem commands are used to return error codes or messages from the voltmeter error queue, control
hardware features and configure communication parameters for the GPIB.


SYSTem:BEEP[:ENABle]

Description:

Set or return the status of the audible keyboard beeper.

Syntax: SYSTem:BEEP[:ENABle]

Argument:

= { 0, 1, OFF, ON }

SYSTem:BEEP:IMMediate

Description:

Causes the system to emit an audible beep. Command only; does not respond to a query.

Syntax: SYSTem:BEEP:IMMediate

Argument: None

SYSTem:COMMunicate:GPIB:ADDRess

Description:

Set or return the GPIB bus address.

Syntax: SYSTem:COMMunicate:GPIB:ADDRess

Argument:

= 1 to 30

SYSTem:COMMunicate:GPIB:EOI

Description:

Set or return the status of the GPIB EOI .

Syntax: SYSTem:COMMunicate:GPIB:EOI

Argument:

= { 0, 1, OFF, ON }

SYSTem:COMMunicate:GPIB:LISTen

Description:

Set or return the GPIB Listen end-of-string terminator .

Syntax: SYSTem:COMMunicate:GPIB:LISTen

Argument:

= {LF, CR, CRLF, EOIONLY}