Boonton 9240 series rf voltmeter, Getting started 3-16 – Boonton 9240 RF Voltmeter User Manual
Page 42
Boonton 9240 Series RF Voltmeter
R T N
M O D E L
> 5 1 0 7 5
S E R
I A L
# 2 9 3 9 1
D A T E
0 8 2 3 1 0
U P S C A L E
0 : 5 0 0 0
D O W N S C A L E 0 : 0 0 0 0
M I
D S C A L E
0 : 0 0 0 0
Figure 3-14. Sensor Edit Data Menu Display
The parameter list will show TBLn (where n = 1, 2, 3, 4) when a serial number has not been entered for the corresponding
internal table. For example, TBL3 will be displayed if the serial number has not been previously entered for internal table 3. In
addition, the parameter list will show ADPTn (where n = 1, 2) if a serial number has not been entered for the table contained
within the sensor data adapter. For example, ADPT2 is displayed when the serial number has not been previously entered for
sensor data adapter 2.
To edit the voltage probe calibration data, move the cursor to the EDIT DATA function and press
voltage probe serial numbers until the desired selection is displayed. Press
depress the
Table 3-7 Sensor Edit Data Menu Functions
Function Description
Parameters
Defaults
RTN
Returns the instrument to the
n/a
n/a
MODEL
Voltage probe model number
0 to 99999
0
SERIAL #
Voltage probe serial number
0 to 99999
0
DATE
Calibration date
MM/DD/YY
01/01/01
Where:
MM = 01 to 12
DD = 01 to 31
YY = 00 to 99
UPSCALE
Upscale linearity factors
Range : Factor
5000
[0 to 6] : [0 to 9999]
DOWNSCALE
Downscale linearity factors
Range : Factor
0
[0 to 6] : [-999 to 999]
MIDSCALE
Midscale linearity factors
Range : Factor
0
[0 to 6] : [-999 to 999]
Getting Started
3-16