beautypg.com
Boonton 9240 Series RF Voltmeter
8. Use the key to move the "<" cursor to SWITCHES and press . Press each
front panel key, avoiding
9. Use the key to select RECORDER and press . This test will sequentially
send a DC voltage in 1 volt steps to the recorder output BNC connector on the rear panel. The test will continue until
10. Press
11. Press the key and verify that the RF Voltage Probe serial number(s) appear under the
channel heading(s). An active channel with no sensor installed will report a table number.
12. Press the key and verify that the filter time and number of samples appear for each active
channel.
13. Press the
channel. Verify the ZERO operation completes successfully.
14. Repeat steps 13 for channel 2 if installed.
15. Connect a GPIB controller to the Model 9240. Verify that the instrument can be addressed to Listen at
its IEEE bus address, and set to Remote. The display must show the correct status on the bottom line of the display. For message passing, the line terminators for the controller and the Model 9240 must be compatible for both Listen and Talk. Use
BOONTON
ELECTRONICS,
9242,
11002,
20110111
16. Connect a dumb terminal or PC serial terminal to the Model 9240. Use a null modem if the terminal is
wired as DCE. For message communication to take place, the parameters of the serial connection and message strings must agree between the terminal and the Model 9240. Use
Installation
2-4