beautypg.com

Boonton Power Sensor User Manual

Page 9

background image

Power sensors have response variations (with respect to the reference frequency) at
high frequencies. Calibration factors ranging from ± 3 dB are entered into the
instrument memories at the desired frequencies. Generally, calibration factors are
within ±0.5 dB. These calibration factors must be traceable to the National Institute
of Standards Technology (NIST) to be meaningful. This is accomplished by sending
a standard power sensor (Thermocouple type) to NIST or a certified calibration house
and comparing this standard sensor against each production sensor. The predominant
error term is the uncertainty of the reference sensor, which is typically 2% to 6%,
depending on the frequency. Refer to Figure 1-3. Calibration Factor Traceability.

NIST

Golden Gate

Calibration Labs

Standard

Sensors

Scalar

Network Analyzer

Sensor

Calibration Factors &

SWR

Figure 1-3. Calibration Factor Traceability

4

Power Sensor Manual