Boonton Power Sensor User Manual
Page 30

Table 3-2. Peak Power Sensor Calibration Factor Uncertainty (con't.)
Models 57318, 57340, 57518, 57540, 58318, 59318
Freq
Model
57318
57340
57518
57540
58318
(1)
59318
(1)
GHz
%
% RSS
%
% RSS
%
% RSS
%
% RSS
%
% RSS
%
% RSS
DUAL DIODE PEAK POWER SENSORS
0.5
1.6
1.1
2.5
1.7
1.6
1.0
2.3
1.5
3.4
1.4
3.4
1.4
1
1.7
1.1
1.9
1.3
1.7
1.1
1.7
1.1
3.4
1.4
3.4
1.4
2
2.0
1.3
2.0
1.3
2.1
1.5
1.8
1.2
3.4
1.4
3.4
1.4
3
2.1
1.5
2.2
1.5
2.2
1.5
2.0
1.3
3.6
1.4
3.6
1.4
4
2.0
1.3
2.4
1.7
2.0
1.3
2.2
1.5
3.8
1.5
3.8
1.5
5
2.2
1.5
2.5
1.8
2.4
1.7
2.3
1.5
3.9
1.5
3.9
1.5
6
2.3
1.7
2.7
2.0
2.2
1.5
2.4
1.6
4.0
1.5
4.0
1.5
7
1.9
1.4
3.0
2.3
1.6
1.1
2.7
1.9
4.0
1.5
4.0
1.5
8
1.9
1.2
3.2
2.6
1.7
1.1
2.9
2.2
4.2
1.6
4.2
1.6
9
2.5
1.7
4.7
4.4
2.5
1.7
4.2
3.8
3.9
1.5
3.9
1.5
10
3.2
2.4
4.8
4.7
3.1
2.2
4.4
4.1
3.7
1.5
3.7
1.5
11
3.6
2.6
4.9
5.0
3.3
2.5
4.5
4.4
4.3
1.6
4.3
1.6
12
3.1
2.2
4.9
5.0
3.3
2.4
4.5
4.4
4.6
1.7
4.6
1.7
13
2.8
1.9
5.1
5.0
3.2
2.5
4.7
4.4
4.1
1.6
4.1
1.6
14
3.5
2.6
5.3
5.0
3.5
2.7
5.0
4.5
3.9
1.5
3.9
1.5
15
3.8
2.8
5.1
4.6
3.4
2.5
5.0
4.4
4.3
1.6
4.3
1.6
16
3.8
2.8
4.9
4.1
4.1
3.0
4.8
4.0
4.5
1.7
4.5
1.7
17
3.2
2.4
4.6
3.7
3.3
2.5
4.7
3.7
4.1
1.6
4.1
1.6
18
3.5
2.5
4.5
3.5
3.4
2.4
4.7
3.7
4.2
1.6
4.2
1.6
19
4.7
3.6
5.2
4.2
20
4.9
3.7
5.4
4.4
21
5.3
4.2
5.7
4.6
22
5.9
4.6
6.0
4.8
23
6.1
4.8
6.3
5.0
24
6.3
4.8
6.6
5.2
25
6.3
4.8
6.3
4.9
26
6.5
5.0
6.6
5.2
26.5
6.9
5.1
7.1
5.2
27
7.4
5.6
7.6
5.8
28
6.6
5.1
6.8
5.3
29
6.4
4.8
6.7
5.1
30
6.3
4.7
6.7
5.0
31
6.5
4.9
6.8
5.1
32
6.9
5.3
6.7
5.0
33
7.3
5.6
6.7
5.0
34
6.6
5.1
6.1
4.7
35
6.5
5.3
5.9
4.6
36
5.7
4.7
5.5
4.5
37
4.9
4.3
5.3
4.6
38
5.6
4.7
6.4
5.7
39
6.4
5.4
8.0
7.5
40
6.6
5.6
8.4
7.9
NOTES: 1) Uncertainty derived in part from the sensor SWR specification applied to a Tegam test system.
Denotes legacy sensors. For reference only. Not for new designs.
Power Sensor Manual
25