Section 3, Power sensor uncertainty factors – Boonton Power Sensor User Manual
Page 22

3
Power Sensor Uncertainty Factors
The uncertainty factors, as a function of frequency for the Diode and Thermocouple,
Peak and Waveguide sensors, are listed in Tables 3-1, 3-2 and 3-3 respectively.
These values represent typical results based on factory test data unless otherwise noted.
The percent (%) column is the sum of all test system uncertainties including mismatch
uncertainties, the uncertainty of the standard sensor and transfer uncertainty which is
traceable to NIST ( National Institute of Standards Technology ). The probable
uncertainty ( % RSS ) is derived by the square root of the sum of the individual
uncertainties squared. % RSS is expressed with a coverage factor of 2 yielding a 95%
confidence level.
Table 3-1. Diode and Thermocouple Power Sensor Calibration Factor Uncertainty
Models 51011(4B), 51011-EMC, 51012(4C), 51013(4E), 51015(5E), 51033(6E)
Freq
Model
(Alias)
51011
51011-EMC
51012
51013
51015
51033
(4B)
(EMC)
(4C)
(4E)
(5E)
(6E)
GHz
%
% RSS
%
% RSS
%
% RSS
%
% RSS
%
% RSS
%
% RSS
WIDE DYNAMIC RANGE DUAL DIODE SENSORS
0.03
1.9
1.1
1.9
1.1
1.8
1.0
2.0
1.1
2.1
1.2
2.0
1.1
0.1
1.7
0.9
1.7
1.0
0.3
1.6
0.9
0.5
1.6
0.9
2.0
1.1
1
1.7
0.9
1.8
1.0
2.3
1.4
1.7
1.0
1.9
1.0
1.7
0.9
1.5
2.4
1.5
2
1.9
1.1
2.1
1.2
2.4
1.4
1.9
1.1
1.9
1.0
1.8
1.0
3
2.0
1.1
2.4
1.4
2.0
1.2
2.2
1.2
1.9
1.0
4
2.1
1.2
2.6
1.6
2.1
1.2
2.3
1.2
1.9
1.1
5
2.2
1.2
2.8
1.7
2.4
1.4
2.0
1.1
2.0
1.1
6
2.5
1.5
3.1
2.2
2.5
1.6
1.9
1.1
1.9
1.0
7
2.5
1.7
3.2
2.5
2.6
1.9
2.0
1.1
1.7
1.0
8
3.0
2.2
3.7
3.1
3.1
2.3
2.2
1.3
2.0
1.1
9
4.9
4.1
5.3
4.6
2.8
1.7
2.8
1.7
10
5.8
4.8
6.1
5.3
3.4
2.3
3.2
2.1
11
6.1
5.2
6.4
5.5
4.2
2.9
3.3
2.3
12
6.3
5.6
6.3
5.7
3.4
2.2
3.2
2.0
13
6.5
6.3
3.7
2.6
3.4
2.2
14
6.6
6.0
4.0
2.7
3.6
2.3
15
7.7
7.2
3.8
2.6
3.2
2.2
16
7.1
6.4
3.7
2.4
3.3
2.2
17
6.7
6.7
3.5
2.3
2.7
1.5
18
6.4
5.7
4.4
3.1
3.6
2.2
Power Sensor Manual