beautypg.com

Record hours of operation & software version – AirSep Eclipse 5 20631679 User Manual

Page 51

background image

Personal Ambulatory Oxygen System Provider Technical Manual

51

PN 20631679 Rev C

RECORD HOuRS OF OPERATION & SOFTWARE VERSION

To help maintain the Eclipse 5, you may obtain the Total Hours of Operation and software version numbers for the Control
Board and the Power Manager Systems by following the steps below.

Alarm Code

Pulse Sensitivity

Hours of

Operation

9-Volt Battery

Status (Volts)

Rise Time

Control Board

Software Revision

Power Manager

Board Software

Revision

NOTE:

The display accrues to “99,999”

hours, then rolls over to “00,000”.

• Alarm Code (ALRM) - Displays most recent
alarm code. Additional alarm codes
will also be displayed, if present, by
continuing to press the + key.
• Pulse Sensitivity (PS)
• Hours of Operation (HRS)
• 9-volt Battery Status (9V)
• Rise Time (BOL)
• Control Board Software Revision
• Power Manager Software Revision


- If connected to the AC or DC Power Supplies, press
the “No Smoking” icon on the control panel and
advance to the following display mode for each time
the icon is depressed:

- If external power is not available and the power
cartridge is installed, start by pressing the ON/OFF key
for 1 second.

- If no key press is sensed within 5 seconds, the Eclipse
5 will drop out of Provider Mode.

Data Output Form Example

Log the Results