Record hours of operation & software version – AirSep Eclipse 5 20631679 User Manual
Page 51
Personal Ambulatory Oxygen System Provider Technical Manual
51
PN 20631679 Rev C
RECORD HOuRS OF OPERATION & SOFTWARE VERSION
To help maintain the Eclipse 5, you may obtain the Total Hours of Operation and software version numbers for the Control
Board and the Power Manager Systems by following the steps below.
Alarm Code
Pulse Sensitivity
Hours of
Operation
9-Volt Battery
Status (Volts)
Rise Time
Control Board
Software Revision
Power Manager
Board Software
Revision
NOTE:
The display accrues to “99,999”
hours, then rolls over to “00,000”.
• Alarm Code (ALRM) - Displays most recent
alarm code. Additional alarm codes
will also be displayed, if present, by
continuing to press the + key.
• Pulse Sensitivity (PS)
• Hours of Operation (HRS)
• 9-volt Battery Status (9V)
• Rise Time (BOL)
• Control Board Software Revision
• Power Manager Software Revision
- If connected to the AC or DC Power Supplies, press
the “No Smoking” icon on the control panel and
advance to the following display mode for each time
the icon is depressed:
- If external power is not available and the power
cartridge is installed, start by pressing the ON/OFF key
for 1 second.
- If no key press is sensed within 5 seconds, the Eclipse
5 will drop out of Provider Mode.
Data Output Form Example
Log the Results