Personal Ambulatory Oxygen System Provider Technical Manual

12

PN 20631679 Rev C

Recommended separation distances

between portable and mobile RF communications equipment and the Eclipse 5

The Eclipse 5 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The cus-

tomer or the user of the Eclipse 5 can help prevent electromagnetic interference by maintaining a minimum distance between

portable and mobile RF communications equipment (transmitters) and the Eclipse 5 as recommended below, according to the

maximum output power of the communications equipment.

Rated maximum out-

put power of transmit-

ter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2,5 GHz

0,01

0.12

0.12

0.23

0,1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can

be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of

the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection

from structures, objects and people.

P

d

2

.

1

P

d

2

.

1

P

d

3

.

2