Personal Ambulatory Oxygen System Provider Technical Manual
50
PN 20631679 Rev C
ASSEMBLy AND ALARM VERIFICATION TESTS
To ensure proper assembly and functionality of the Eclipse 5 after it has been reassembled, the following steps should be
followed.
1. Install the Power Cartridge into the Power Cartridge compartment of the Eclipse 5. Plug the AC Power Supply into the
wall outlet and connect it to the External Power Connector of the Eclipse 5.
2. Press the ON button and set the Eclipse 5 to 2 LPM. At initial start-up Eclipse 5 units light the Green, Yellow and Red
LED’s accompanied by a beep. As concentration increases to ~70%, the Red LED turns off and the Yellow and Green
remain on. When the concentration reaches normal operating range (above 85%), the Yellow LED will turn off. The
Green LED will stay lit indicating normal operation. Table 3 shows the normal start-up operating conditions for Eclipse
5 Oxygen Systems. If LED is not green within 7 minutes there may be a problem with the Eclipse 5.
Green
Indicator
Yellow
Indicator
Red Indicator
Audible
Alarm
Alarm code
Operating Condition
Off
Off
Blinking
Beeps
004
Purity < 70%
Off
Blinking
Off
Off
008
Purity between 70% and
85%
On
Off
Off
Off
000
Normal Operating Condition
3. Verify that the Power Cartridge is charging as indicated by the Power Cartridge Status Gauge is moving in waterfall
fashion (scrolling from bottom to top). If the Power Cartridge Status Gauge is not moving, verify that the Power
Cartridge is properly engaged.
4. Check the Control Panel by pressing each of the Increase/Decrease, Pulse Dose Mode and No Smoking buttons and
observe that the Eclipse 5 buttons function normally.
5. Press the Delivery Mode button and set the Eclipse 5 to the patient’s normal Pulse Dose setting. Confirm that without
breathing from the unit, after 15 seconds of no breath detected, the system changes automatically to Continuous
Flow Mode. The green Delivery Mode light is blinking fast, indicating the system is delivering a Continuous Flow. After
another 15 seconds, the system stops delivering Continuous Flow and waits 15 seconds trying to detect a breath. The
cycle continues until a breath is detected.
6. After 5 minutes of operation, block the Cannula Fitting Outlet with your finger for 2 minutes and confirm that the
visual and audible alarm occurs. Unblock the Cannula Outlet Port and let it stabilize until the alarm stops.
7. Disconnect the AC Power Supply and allow the Eclipse 5 to run for about one minute. Set the Eclipse 5 to the
patient’s normal Continuous Mode setting. Confirm that the Power Cartridge Status Gauge is illuminated and External
Power Present Indicator is off.
8. Remove the Power Cartridge and verify that the Eclipse 5 alarms and red Alarm Indicator is flashing and the alarm
sounds. After about 5 seconds, re-install the Power Cartridge and connect the AC Adapter and observe that the
Eclipse 5 automatically restarts. Confirm that the External Power Present Indicator is illuminated.
9. Turn off Eclipse 5 and unplug the AC Power Supply.
10. Record results, initial and date the Service and Maintenance Record.
11. EDAT may be utilized to record Eclipse 5 device status and dates of service.